What are the parameters required to operate the Apollo according to ISO 105-B02:2014?
This configuration uses the SolarSens radiometer with 300 - 400nm sensor to control the irradiance at 42 W/m2 (300-400nm).

40% using the Red Azoic Fabric and Blue Wool Reference No. 5.
See section 8 of ISO 105-B02:2014 for more details.