- INSTRUMENTS
- Orbitor
-
INSTRUMENTS
- Titan 4, 5 and 10
- TruBurst
- ThermaPlate
- Colour Assessment Cabinets
- Sample Cutters
- Titan 1 2 3
- FlexiBurn
- TruFade
- ElastAbrasion
- Spray Rating Tester
- Martindale
- Orbitor
- GyroWash
- Perspirometer
- DynaWash & DynaWash Duo
- SnagPod
- FlexiFrame
- CrockMaster
- ElmaTear
- Impulse
- Wascator
- ProView
- ProMace
- AccuDry
- Crease Recovery Angle Tester
- Apollo
- Wrinkle Recovery Tester
- Titan
- DynaWash
-
OPERATOR'S GUIDES
- Memmert Incubator IN30
- GyroWash
- CrockMaster
- Spray Rating Tester
- ElmaTear
- Titan
- Martindale 900 Series
- AquAbrasion 1819
- TruFade
- Martindale 1600 Series
- SnagPod
- Perspirometer
- Wascator
- TruBurst
- DynaWash & DynaWash Duo
- Orbitor
- Crease Recovery Angle Tester
- Martindale 1300 Series
- Yellowing Test Kit
- Tautex
- ElastAbrasion
- ThermaPlate
- FlexiBurn
- Incubator / Oven
- AccuDry
- Martindale model 103
- Wrinkle Recovery Tester
- Drape Tester
- ProMace
- Sample Cutters
- Sock Abrasion Kit
- Impulse
- ProView
- FlexiFrame
- General
- INSTALLATION GUIDE
-
TEST MATERIALS
-
EU DECLARATION OF CONFORMITY
-
MATERIAL SAFETY DATA SHEETS
-
GENERAL
-
Filters
Are James Heal instrument liners ready to use?
Orbitor
Run in before use:
- Cork Liners for pilling
- Cork Liners for snagging
These should both be run in using four empty polyurethane tubes for 200 hours at 60rpm.
The setting for running in cork liners is included in the software on Orbitor.
Thorough cleaning must be carried out to remove cork dust before testing commences.
Random Tumble Pilling Tester
Ready to use:
- Cork Liner for Impulse (393-527)
- Cork Liners – Extra Long (393-533)
Require preparation:
- Neoprene Liners (758–566)
Neoprene liners need to be cleaned on both sides with a paste made with IEC A Detergent to remove any residue from the manufacturing process.
After cleaning they should be run in on both sides using 3 specimens of the standard in-house reference fabric.
Detailed instructions of these procedures can be found in annex A of ISO 12945-3.